Measured variables and measuring ranges
|
Connection type mV:
pH: 0.00 … 14.00
ORP potential: -1500 … +1500 mV
Type of connection mA (amperometric measured variables, measuring ranges corresponding to sensors):
Chlorine
Chlorine dioxide
Chlorite
Bromine
Ozone
Hydrogen peroxide (PER 1 and PEROX H-3E)
Hydrogen peroxide (PEROX H2.10 P with PEROX transducer V2 order. no. 1047979)
Peracetic acid
Dissolved oxygen
Type of connection mA (potentiometic measured variables, measuring ranges corresponding to transmitters):
pH
ORP potential
Fluoride:
via module VA and function upgrade package 3 and 4
Conductivity mA via sensor CCT 1-mA-20 mS/cm
Temperature:
via Pt 100/Pt 1000, measuring range 0 … 150 °C
|
|
Resolution
|
pH: 0.01
ORP voltage: 1 mV
temperature: 0.1 °C
amperometric analysis (chlorine etc.): 0.001/0.01 ppm, 0.01 vol.%, 0.1 vol.%
|
|
Accuracy
|
0.3% based on the full-scale reading
|
|
Measurement input
|
pH/ORP (input resistance > 0.5 x 1012 Ω)
|
|
Temperature compensation
|
Pt 100/Pt 1000 for pH, chlorine dioxide (CDP) sensor and fluoride
|
|
Correction range temp.
|
0…100 °C
|
|
pH compensation range for chlorine
|
Sensor CLE 3 and CLE 3.1: 6.5 … 8.5, sensor CBR: 6.5 … 9.5
|
|
Disturbance signals
|
Flow via 0/4 … 20 mA signal or contact water meter, 1 – 500 Hz. The multiplicative interference variable can influence all channels, while the additive interference variable only influences one channel.
|
|
Control characteristic
|
P/PID control
|
|
Control
|
2 or 3 bidirectional controls
|
|
Analogue outputs
|
2 (3) x 0/4 … 20 mA electrically isolated, max. load 450 Ω, range and assignment (measured, correction, control variable) can be set
|
|
Control outputs
|
2 (4) pulse frequency outputs for controlling metering pumps
2 relays (limit value or pulse length control)
|
|
Alarm relay
|
250 V ~3 A, 700 VA contact type, changeover
|
|
Digital control inputs
|
4 (7) as a remote control input for the functions pause control/sample water fault, parameter set switch-over, level monitoring of chemical tanks
|
|
Electrical Connection
|
100 – 230 V, 50/60 Hz, 25 VA, optional 24 V DC
|
|
Field bus connection
|
PROFIBUS® DP, Modbus RTU, PROFINET
|
|
Ambient temperature
|
0 … 50 °C (for use indoors or with a protective enclosure)
|
|
Enclosure rating
|
Wall-mounted: IP 66 and IP 67 (NEMA 4X)
Mounted in control cabinet: IP 54 for control cabinet door
|
|
Tests and approvals
|
CE and MET (corresponding to UL as per IEC 61010)
|
|
Housing material
|
PC with flameproofing equipment
|
|
Dimensions
|
250 x 220 x 122 mm (WxHxD)
|
|
Weight
|
1.3 kg
|